Hyppää sisältöön
    • Suomeksi
    • In English
Trepo
  • Suomeksi
  • In English
  • Kirjaudu
Näytä viite 
  •   Etusivu
  • Trepo
  • Opinnäytteet - ylempi korkeakoulututkinto
  • Näytä viite
  •   Etusivu
  • Trepo
  • Opinnäytteet - ylempi korkeakoulututkinto
  • Näytä viite
JavaScript is disabled for your browser. Some features of this site may not work without it.

Cone-Beam Computed Tomography Examinations of the Head and Neck Region in Finland: Indications and Patient Radiation Dose

Turnbull-Smith, Sonja Benita (2016)

 
Avaa tiedosto
Master of Science thesis (2.679Mt)
Lataukset: 



Turnbull-Smith, Sonja Benita
2016

Biotekniikan koulutusohjelma
Luonnontieteiden tiedekunta - Faculty of Natural Sciences
This publication is copyrighted. You may download, display and print it for Your own personal use. Commercial use is prohibited.
Hyväksymispäivämäärä
2016-09-07
Näytä kaikki kuvailutiedot
Julkaisun pysyvä osoite on
https://urn.fi/URN:NBN:fi:tty-201608264465
Tiivistelmä
Diagnostic reference levels (DRLs) are pre-determined dose levels employed for patient dose management in diagnostic radiology. Cone-beam computed tomography (CBCT) has gained significant popularity in the imaging of the head and neck region eliciting the need for DRLs for CBCT examinations.
The main objective of the study was to collect dose data for the establishment of indication-based national DRLs for CBCT examinations of the head and neck region in Finland as well as to form proposals for the prospective DRLs. In addition, examination frequencies, the use of optimised imaging parameters and the effect of manufacturer and model of the CBCT device on dose were studied.
The study was conducted as a nation-wide dose survey among the Finnish facilities conducting CBCT examinations of the head and neck region. The response rate was 56% in terms of registered devices and 57% in terms of listed places of use.
DRL proposals were made for seven indications. The values for the proposals were derived from the 75th percentile values of the indication-specific dose distributions. Very wide dose ranges were observed within all seven indications. The most common indication was pre-surgical imaging of implant treatments. A significant portion of the respondents had used default factory settings instead of optimising the imaging parameters. No particular manufacturer or device model stood out as producing exceptionally high or low doses regardless of the user. The Finnish DRLs for CBCT examinations will be issued based on the results of this study.
Kokoelmat
  • Opinnäytteet - ylempi korkeakoulututkinto [40600]
Kalevantie 5
PL 617
33014 Tampereen yliopisto
oa[@]tuni.fi | Tietosuoja | Saavutettavuusseloste
 

 

Selaa kokoelmaa

TekijätNimekkeetTiedekunta (2019 -)Tiedekunta (- 2018)Tutkinto-ohjelmat ja opintosuunnatAvainsanatJulkaisuajatKokoelmat

Omat tiedot

Kirjaudu sisäänRekisteröidy
Kalevantie 5
PL 617
33014 Tampereen yliopisto
oa[@]tuni.fi | Tietosuoja | Saavutettavuusseloste