Turning up the current – optimizing stimulation parameters with vagus nerve stimulation in a nationwide Norwegian cohort (The NOR-current study)
Kostov, Konstantin Hrisimirov; Kostov, Hrisimir; Larsson, Pål Gunnar; Henning, Oliver; Egge, Arild; Lossius, Morten Ingvar; Peltola, Jukka (2025-07)
Kostov, Konstantin Hrisimirov
Kostov, Hrisimir
Larsson, Pål Gunnar
Henning, Oliver
Egge, Arild
Lossius, Morten Ingvar
Peltola, Jukka
07 / 2025
BRAIN STIMULATION
Julkaisun pysyvä osoite on
https://urn.fi/URN:NBN:fi:tuni-202509038933
https://urn.fi/URN:NBN:fi:tuni-202509038933
Kuvaus
Peer reviewed
Tiivistelmä
Background: Despite widespread use of vagus nerve stimulation (VNS), evidence regarding optimal stimulation titration and dosing remains limited. Significant inter-practice variability and deviation from guidelines, results in sub-optimal dosing. Objective: Identify optimal stimulation parameters and establish target dosing for VNS therapy in epilepsy. Methods: We linked long-term outcomes with dosing parameters in 415 patients from a VNS registry (median follow-up = 79.0 months). A generalized linear mixed model (GLMM) assessed relationships between stimulation parameters (Output Current (OC), Frequency, Pulse Width (PW), Duty cycle [DC]) and very good clinical response (≥75 % seizure frequency reduction from baseline). Results: Output current was the strongest predictor of seizure reduction (p = 0.001). Increasing OC by 1 mA more than doubled the probability of ≥75 % seizure reduction, peaking at 2.70 mA. Greatest likelihood of ≥50 % seizure reduction was at 2.1 mA. Post-hoc analysis supported GLMM findings, showing significantly more patients (46.3 %) receiving >2.0 mA OC achieved ≥75 % seizure reduction compared with 30.7 % (p = 0.01) among patients using lower OC. Although changes in DC were not significantly associated with seizure reduction, post-hoc analysis suggested benefits from other duty cycles. Standard 250 μs PW and 20 Hz frequency were associated with best effect. Longer follow-up improved seizure control (p < 0.001); intellectual disability negatively impacted response (p = 0.005). Seizure-free patients had significantly lower total charge (122.4 vs. 250.1 mC, p = 0.014) than patients with persistent seizures. Conclusion: Output current should be the primary focus when titrating VNS patients to optimal dose. We suggest a target range of 1.5–2.25 mA with 250 μs PW and 20 Hz frequency. Higher OC should be attempted if tolerated.
Kokoelmat
- TUNICRIS-julkaisut [22172]