Median 10-year whole blood metal ion levels and clinical outcome of ReCap-M2a-Magnum metal-on-metal total hip arthroplasty

Pietiläinen, Sakari; Linnovaara, Axel; Venäläinen, Mikko S.; Mäntymäki, Heikki; Laaksonen, Inari; Lankinen, Petteri; Mäkelä, Keijo T. (2022-04)

 
Avaa tiedosto
16278_PIETIL_INEN_D.pdf (838.6Kt)
Lataukset: 



Pietiläinen, Sakari
Linnovaara, Axel
Venäläinen, Mikko S.
Mäntymäki, Heikki
Laaksonen, Inari
Lankinen, Petteri
Mäkelä, Keijo T.
04 / 2022

Acta orthopaedica
doi:10.2340/17453674.2022.2510
Näytä kaikki kuvailutiedot
Julkaisun pysyvä osoite on
https://urn.fi/URN:NBN:fi:tuni-202206035472

Kuvaus

Peer reviewed
Tiivistelmä
<p>BACKGROUND AND PURPOSE: We have previously reported that the whole blood (WB) chromium (Cr) and cobalt (Co) ion levels decrease in the short term after ReCap-M2a-Magnum large-diameter head (LDH) metal-on-metal (MoM) total hip arthroplasty (THA). This study reports long-term metal ion levels and clinical outcomes after ReCap-Magnum THA. PATIENTS AND METHODS: ReCap-M2a-Magnum LDH THA was used in 1,450 patients in our hospital district from 2005 to 2012. Median follow-up time was 10 years. 991 patients had 2 or more metal ion measurements. The median measurement interval was 4 years. Individual metal ion change was assessed using logarithmic metal ion values in a random coefficient model. Kaplan-Meier survival estimates were calculated for revision surgery for any reason for revision, and separately for metal-related adverse events (metal ions above safe upper limit [SUL], revision due to ARMD, or pseudotumor). RESULTS: Geometric mean of Cr decreased from 1.8 ppb (geometric standard deviation [GSD] 1.8) to 1.0 ppb (GSD 2.8, p < 0.001). The Co levels decreased from 1.7 ppb (GSD 2.4) to 1.4 ppb (GSD 2.8, p < 0.001). The hip-specific survival was 85% for revision due to any reason at 14 years and the hip-specific survival for any metal-related adverse event was 69% at 14 years. INTERPRETATION: WB Cr and Co levels continued to decrease in the long-term follow-up of ReCap-M2a-Magnum THA patients. The amount of metal-related adverse events was rather high, but revision surgery was seldom required. We suggest that after 10 years from the implantation a 5-year measurement interval may be sufficient for asymptomatic ReCap-M2a-Magnum patients.</p>
Kokoelmat
  • TUNICRIS-julkaisut [20683]
Kalevantie 5
PL 617
33014 Tampereen yliopisto
oa[@]tuni.fi | Tietosuoja | Saavutettavuusseloste
 

 

Selaa kokoelmaa

TekijätNimekkeetTiedekunta (2019 -)Tiedekunta (- 2018)Tutkinto-ohjelmat ja opintosuunnatAvainsanatJulkaisuajatKokoelmat

Omat tiedot

Kirjaudu sisäänRekisteröidy
Kalevantie 5
PL 617
33014 Tampereen yliopisto
oa[@]tuni.fi | Tietosuoja | Saavutettavuusseloste