Comparison of three differently shaped ROIs in free breathing breast radiotherapy setup using surface guidance with AlignRT

Laaksomaa, Marko; Moser, Torsten; Kritz, Julia; Pynnönen, Kiira; Rossi, Maija (2021)

 
Avaa tiedosto
75829_310220_1_PB.pdf (661.2Kt)
Lataukset: 



Laaksomaa, Marko
Moser, Torsten
Kritz, Julia
Pynnönen, Kiira
Rossi, Maija
2021

Reports of Practical Oncology and Radiotherapy
doi:10.5603/RPOR.a2021.0062
Näytä kaikki kuvailutiedot
Julkaisun pysyvä osoite on
https://urn.fi/URN:NBN:fi:tuni-202112058915

Kuvaus

Peer reviewed
Tiivistelmä
<p>Background: Setup accuracy within adjuvant radiotherapy of breast cancer treated in free breathing is well studied, but a comparison of the typical regions of interest (ROI) used in surface guided radiation therapy (SGRT) does not exist. The aim of this study was to estimate the setup accuracy obtained with differently shaped ROIs in SGRT. Materials and methods: A total of 573 orthogonal image pairs were analyzed from free breathing breast patients in two groups: positioning using AlignRT® surface guidance system (Group A, n = 20), and setup using conventional laser and tattoo setup (Group L, n = 20). For SGRT, three different setup ROIs were used: a Breast-shaped, O-shaped and T-shaped (B-O and T-ROI). We evaluated the isocenter-, rotation-, pitch and arm position accuracy and residual errors for the chest wall and shoulder joint in kV orthogonal and tangential setup images with laser or SGRT-based setup. Results: Less isocenter variance was found in Group A than in Group L. Rotations and posture errors were larger in group L than in Group A (p ≤ 0.05). Rotation error was smaller with T-shaped ROI than with O-or B-shape (p = 0.01–0.04). Conclusion: Setup with AlignRT® improves reproducibility compared to laser setup. Between the different ROI shapes only small differences were found in the patient posture or the isocenter position in the images. The T-ROI is recommended to set up the chest wall bony structure and an additional B-ROI may be used to fine-tune the soft tissue accuracy</p>
Kokoelmat
  • TUNICRIS-julkaisut [20683]
Kalevantie 5
PL 617
33014 Tampereen yliopisto
oa[@]tuni.fi | Tietosuoja | Saavutettavuusseloste
 

 

Selaa kokoelmaa

TekijätNimekkeetTiedekunta (2019 -)Tiedekunta (- 2018)Tutkinto-ohjelmat ja opintosuunnatAvainsanatJulkaisuajatKokoelmat

Omat tiedot

Kirjaudu sisäänRekisteröidy
Kalevantie 5
PL 617
33014 Tampereen yliopisto
oa[@]tuni.fi | Tietosuoja | Saavutettavuusseloste