Performance of thin Vivacit-E hip liners: no relevant wear during ex vivo testing at high acetabular inclination angle

Saikko, Vesa; Neuvonen, Perttu; Reito, Aleksi (2022-12-02)

 
Avaa tiedosto
Acta_Orthopaedica_2022_93_901_905.pdf (1.654Mt)
Lataukset: 



Saikko, Vesa
Neuvonen, Perttu
Reito, Aleksi
02.12.2022

ACTA ORTHOPAEDICA
doi:10.2340/17453674.2022.5359
Näytä kaikki kuvailutiedot
Julkaisun pysyvä osoite on
https://urn.fi/URN:NBN:fi:tuni-202212199327

Kuvaus

Peer reviewed
Tiivistelmä
<p>Background and purpose — There is concern among clinicians regarding the performance of thin, highly cross-linked polyethylene acetabular liners at high inclination angles that cause edge contact and high contact stresses. We studied ex vivo wear performance of thin, vitamin-E grafted, highly cross-linked polyethylene (Vivacit-E) liners in rela-tion to high acetabular inclination angle. Materials and methods — Wear of Vivacit-E acetabu-lar liners (thickness 4.0–5.7 mm) was studied with a vali-dated hip joint simulator at 2 different acetabular inclination angles, 40° (optimal) and 65° (high). The test simulated walking. Wear was evaluated gravimetrically and dimen-sionally. Results — At the optimal inclination angle, slight weight gain occurred. At the high angle, the liners lost weight slightly. Due to the minimal weight loss, gravimetric wear rates were difficult to determine. Linear wear was below the detection limit of 0.01 mm. Conclusion — Even with the high acetabular inclination angle, the prostheses performed well in the present test conditions. The wear rates of the liners were estimated to be clearly below the osteolysis threshold of 0.05 mm/year, below which osteolysis is absent. Since the present hip joint simulator has been shown to produce clinically relevant wear, these ex vivo results are likely to reflect the in vivo behavior of the design.</p>
Kokoelmat
  • TUNICRIS-julkaisut [20161]
Kalevantie 5
PL 617
33014 Tampereen yliopisto
oa[@]tuni.fi | Tietosuoja | Saavutettavuusseloste
 

 

Selaa kokoelmaa

TekijätNimekkeetTiedekunta (2019 -)Tiedekunta (- 2018)Tutkinto-ohjelmat ja opintosuunnatAvainsanatJulkaisuajatKokoelmat

Omat tiedot

Kirjaudu sisäänRekisteröidy
Kalevantie 5
PL 617
33014 Tampereen yliopisto
oa[@]tuni.fi | Tietosuoja | Saavutettavuusseloste