Sensitization potential of medical devices detected by in vitro and in vivo methods

Svobodová, Lada; Rucki, Marian; Vlkova, Alena; Kejlova, Kristina; Jírová, Dagmar; Dvorakova, Marketa; Kolarova, Hana; Kandárová, Helena; Pôbiš, Peter; Heinonen, Tuula; Maly, Marek (2021)

 
Avaa tiedosto
2017_PDF_17678_2_10_20210720.pdf (1.016Mt)
Lataukset: 



Svobodová, Lada
Rucki, Marian
Vlkova, Alena
Kejlova, Kristina
Jírová, Dagmar
Dvorakova, Marketa
Kolarova, Hana
Kandárová, Helena
Pôbiš, Peter
Heinonen, Tuula
Maly, Marek
2021

ALTEX
doi:10.14573/altex.2008142
Näytä kaikki kuvailutiedot
Julkaisun pysyvä osoite on
https://urn.fi/URN:NBN:fi:tuni-202108266818

Kuvaus

Peer reviewed
Tiivistelmä
<p>Medical devices must be tested before marketing in accordance with ISO EN 10993-10 in order to avoid skin sensi­tization. This standard predominantly refers to the in vivo test but does not exclude the use of in vitro methods that have been sufficiently technically and scientifically validated for medical device testing. It is foreseen that, due to the complexity of the sensitization endpoint, a combination of several methods will be needed to address all key events occurring in the sensitization process. The objective of this pilot study was to evaluate the sensitization potential of selected medical devices using a combination of in chemico (DPRA, OECD TG 442C) and in vitro (LuSens, OECD TG 442D) methods in comparison with the in vivo (LLNA DA, OECD TG 442A) method and to suggest a possible testing strategy for the safety assessment of medical device extracts. Overall, one of the 42 tested samples exhibited positive results in all employed test methods, while 33 samples were predicted as non-sensitizing in all three performed methods. This study demonstrated good agreement between in vitro and in vivo results regarding non-sensitizing samples; however, some discrepancies in positive classification were recorded. A testing strategy is suggested in which negative results are accepted and any positive results in the in chemico or in vitro tests are followed up with a third in vitro test and evaluated in accordance with the “2 out of 3 approach”. This strategy may reduce and replace animal use for testing the sensitization potential of medical devices.</p>
Kokoelmat
  • TUNICRIS-julkaisut [20709]
Kalevantie 5
PL 617
33014 Tampereen yliopisto
oa[@]tuni.fi | Tietosuoja | Saavutettavuusseloste
 

 

Selaa kokoelmaa

TekijätNimekkeetTiedekunta (2019 -)Tiedekunta (- 2018)Tutkinto-ohjelmat ja opintosuunnatAvainsanatJulkaisuajatKokoelmat

Omat tiedot

Kirjaudu sisäänRekisteröidy
Kalevantie 5
PL 617
33014 Tampereen yliopisto
oa[@]tuni.fi | Tietosuoja | Saavutettavuusseloste