The reliability of radiation dose display of a computed tomography scanner

Forss, V.; Yli-Ollila, H.; Vatanen, J.; Kölhi, P.; Poutanen, V. P.; Palomäki, A. (2021)

 
Avaa tiedosto
the_reliability_of_radiation_2021.pdf (1.172Mt)
Lataukset: 



Forss, V.
Yli-Ollila, H.
Vatanen, J.
Kölhi, P.
Poutanen, V. P.
Palomäki, A.
2021

European Journal of Radiology Open
100345
doi:10.1016/j.ejro.2021.100345
Näytä kaikki kuvailutiedot
Julkaisun pysyvä osoite on
https://urn.fi/URN:NBN:fi:tuni-202104263514

Kuvaus

Peer reviewed
Tiivistelmä
<p>Background: Internationally, the typical allowed difference between the measured radiation dose and dose reported by a computed tomography (CT) scanner is ±20 %. The objective is to describe a method in order to analyse this difference in a CT scanner in the Emergency Department of Kanta-Häme Central Hospital, and to calculate a correction factor for more comparable radiation dose values in further studies. Methods: Ten intra-day radiation dose measurements were performed with undisturbed setting. Measurement reports on differences between measured and displayed dose were gathered from the vendor maintenance and supervising authority over a 12-year period. Additionally, two in-house measurements were made. A total of 18 datapoints were collected, with some differences in measurement settings. Data were also analysed against imaging parameters, ambient air pressure and time to identify trends or associations in the variation of the discrepancy. Results: Measured doses were generally lower than displayed doses. Differences between displayed and measured doses varied between −3.46 and −0.10 %, with a mean of −1.26 % in the intra-day measurements, and between +4.65 and −17.3 %, with a mean of −7.53 % in the long-term data. There were no trends nor connections in the variations. Conclusion: Since the acceptable difference between the radiation dose display and the measured dose is relevant, the average difference for every CT scanner should be determined before radiation dose studies, especially when comparing multiple scanners.</p>
Kokoelmat
  • TUNICRIS-julkaisut [20127]
Kalevantie 5
PL 617
33014 Tampereen yliopisto
oa[@]tuni.fi | Tietosuoja | Saavutettavuusseloste
 

 

Selaa kokoelmaa

TekijätNimekkeetTiedekunta (2019 -)Tiedekunta (- 2018)Tutkinto-ohjelmat ja opintosuunnatAvainsanatJulkaisuajatKokoelmat

Omat tiedot

Kirjaudu sisäänRekisteröidy
Kalevantie 5
PL 617
33014 Tampereen yliopisto
oa[@]tuni.fi | Tietosuoja | Saavutettavuusseloste