Multi-Axis Fatigue Experimentation System of Intramedullary Implants for Femur and Tibia

Kanerva, Mikko; Pärnänen, Tuomas; Jokinen, Jarno; Haaja, Juha; Ritvanen, Antti; Schlenzka, Dietrich (2019-12-02)

 
Avaa tiedosto
Kanerva_et_al_2019_Journal_of_Orthopaedic_Research_.pdf (7.318Mt)
Lataukset: 



Kanerva, Mikko
Pärnänen, Tuomas
Jokinen, Jarno
Haaja, Juha
Ritvanen, Antti
Schlenzka, Dietrich
02.12.2019

Journal of Orthopaedic Research
doi:10.1002/jor.24545
Näytä kaikki kuvailutiedot
Julkaisun pysyvä osoite on
https://urn.fi/URN:NBN:fi:tuni-202001161333

Kuvaus

Peer reviewed
Tiivistelmä
Current designs of leg-lengthening implants have faced serious failures due to inadequacies in the mechanical design. The failure typically is the result of fatigue induced by a combined loading condition with axial and shear components acting in the tubular body of the implant. One of the reasons leading to the failure is improper verification testing for the design of the fatigue limit. The current test standards for pre-clinical design phases of nail implants are relatively straightforward and widely accepted yet cannot produce the three-dimensional stress state representative of the anticipated operation in a patient during the consolidation phase. This work introduces a major improvement toward a method for verifying fatigue life of tubular as well as solid implants under combined torque, axial load, and bending. The report describes a new loading fixture, a calibration method, and compares the qualification results of finite element simulation analyses and experimental measurements during cyclic loading tests. The findings state that the fixture produces controlled multi-axial loadings to study varied osteotomy locations, quasi-static strength and fatigue of intramedullary implants at an intermediate, 2 Hz, cycle rate.
Kokoelmat
  • TUNICRIS-julkaisut [25018]
Kalevantie 5
PL 617
33014 Tampereen yliopisto
oa[@]tuni.fi | Tietosuoja | Saavutettavuusseloste
 

 

Selaa kokoelmaa

TekijätNimekkeetTiedekunta (2019 -)Tiedekunta (- 2018)Tutkinto-ohjelmat ja opintosuunnatAvainsanatJulkaisuajatKokoelmat

Omat tiedot

Kirjaudu sisäänRekisteröidy
Kalevantie 5
PL 617
33014 Tampereen yliopisto
oa[@]tuni.fi | Tietosuoja | Saavutettavuusseloste