Technical Note: Noise models for virtual clinical trials of digital breast tomosynthesis
Borges, Lucas R.; Barufaldi, Bruno; Caron, Renato F.; Bakic, Predrag R.; Foi, Alessandro; Maidment, Andrew D.A.; Vieira, Marcelo A.C. (2019-06-01)
Medical Physics
https://urn.fi/URN:NBN:fi:tuni-202002212296
Kuvaus
Tiivistelmä
Purpose: To investigate the use of an affine-variance noise model, with correlated quantum noise and spatially dependent quantum gain, for the simulation of noise in virtual clinical trials (VCT) of digital breast tomosynthesis (DBT). Methods: Two distinct technologies were considered: an amorphous-selenium (a-Se) detector with direct conversion and a thallium-doped cesium iodide (CsI(Tl)) detector with indirect conversion. A VCT framework was used to generate noise-free projections of a uniform three-dimensional simulated phantom, whose geometry and absorption match those of a polymethyl methacrylate (PMMA) uniform physical phantom. The noise model was then used to generate noisy observations from the simulated noise-free data, while two clinically available DBT units were used to acquire projections of the PMMA physical phantom. Real and simulated projections were then compared using the signal-to-noise ratio (SNR) and normalized noise power spectrum (NNPS). Results: Simulated images reported errors smaller than 4.4% and 7.0% in terms of SNR and NNPS, respectively. These errors are within the expected variation between two clinical units of the same model. The errors increase to 65.8% if uncorrelated models are adopted for the simulation of systems featuring indirect detection. The assumption of spatially independent quantum gain generates errors of 11.2%. Conclusions: The investigated noise model can be used to accurately reproduce the noise found in clinical DBT. The assumption of uncorrelated noise may be adopted if the system features a direct detector with minimal pixel crosstalk.
Kokoelmat
- TUNICRIS-julkaisut [19374]