Possible strategy to use differential mobility spectrometry in real time applications

Anttalainen, Osmo; Puton, J.; Kontunen, Anton; Karjalainen, Markus; Kumpulainen, Pekka; Oksala, Niku; Safaei, Zahra; Roine, Antti (2019)

 
Avaa tiedosto
Anttalainen2019_Article_PossibleStrategyToUseDifferent.pdf (1004.Kt)
Lataukset: 



Anttalainen, Osmo
Puton, J.
Kontunen, Anton
Karjalainen, Markus
Kumpulainen, Pekka
Oksala, Niku
Safaei, Zahra
Roine, Antti
2019

International Journal for Ion Mobility Spectrometry
doi:10.1007/s12127-019-00251-1
Näytä kaikki kuvailutiedot
Julkaisun pysyvä osoite on
https://urn.fi/urn:nbn:fi:tty-201907081950Julkaisun pysyvä osoite on
https://urn.fi/urn:nbn:fi:tuni-202002061892

Kuvaus

Peer reviewed
Tiivistelmä
<p>Differential Mobility Spectrometry (DMS), also called as FAIMS is a variation of atmospheric pressure ion mobility measurement techniques and is capable of providing information about the electric field - mobility dependence of ions. In this method, a combined electric field is used. This field consists of asymmetric oscillating electric field of high intensity and low static field component. Analytical information in DMS is 2-dimensional dependence of ionic current on oscillating field amplitude and the value of static field intensity. The measurement of DMS signal for whole ranges of both variables is time consuming and also generates lot of data. It is a disadvantage of DMS method, which limits the use of this otherwise powerful technology in real time applications that require a response time of few seconds. This paper presents a way to limit measurement time by heuristic knowledge of the properties of the data space and another method based on the concept of Shannon Entropy to find operating parameters satisfying both separation and signal to noise ratio requirements.</p>
Kokoelmat
  • TUNICRIS-julkaisut [22191]
Kalevantie 5
PL 617
33014 Tampereen yliopisto
oa[@]tuni.fi | Tietosuoja | Saavutettavuusseloste
 

 

Selaa kokoelmaa

TekijätNimekkeetTiedekunta (2019 -)Tiedekunta (- 2018)Tutkinto-ohjelmat ja opintosuunnatAvainsanatJulkaisuajatKokoelmat

Omat tiedot

Kirjaudu sisäänRekisteröidy
Kalevantie 5
PL 617
33014 Tampereen yliopisto
oa[@]tuni.fi | Tietosuoja | Saavutettavuusseloste