Vibration transmittance measures sternotomy stability - a preliminary study in human cadavers

Hautalahti, Juha; Joutsen, Atte; Goebeler, S; Luukkaala, T; Khan, J; Hyttinen, J; Laurikka, Jari (2019)

 
Avaa tiedosto
s13019_018_0823_5.pdf (2.095Mt)
Lataukset: 



Hautalahti, Juha
Joutsen, Atte
Goebeler, S
Luukkaala, T
Khan, J
Hyttinen, J
Laurikka, Jari
2019

JOURNAL OF CARDIOTHORACIC SURGERY
2
doi:10.1186/s13019-018-0823-5
Näytä kaikki kuvailutiedot
Julkaisun pysyvä osoite on
https://urn.fi/urn:nbn:fi:tuni-201906252216

Kuvaus

Peer reviewed
Tiivistelmä
<p>Background: Stability is essential for the normal healing of a sternotomy. Mechanical vibration transmittance may provide a new means of early detection of diastasis in the sternotomy and thus enable the prevention of further complications. We sought to confirm that vibration transmittance detects sternal diastasis in human tissue. Methods: Ten adult human cadavers (8 males and 2 females) were used for sternal assessments with a device constructed in-house to measure the transmittance of a vibration stimulus across the median sternotomy at the second, third, and fourth costal cartilage. Intact bone was compared to two fixed bone junctions, namely a stable wire fixation and an unstable wire fixation with a 10 mm wide diastasis mimicking a widely rupturing sternotomy. A generalized Linear Mixed Model with the lme function was used to determine the ability of the vibration transmittance device to differentiate mechanical settings in the sternotomy. Results: The transmitted vibration power was statistically significantly different between the intact chest and stable sternotomy closure, stable and unstable closure, as well as intact and unstable closure (t-values and p-values respectively: t = 6.87, p < 0.001: t = 7.41, p < 0.001: t = 14.3, p < 0.001). The decrease of vibration transmittance from intact to stable at all tested costal levels was 78%, from stable to unstable 58%, and from intact to unstable 91%. The vibration transmittance power was not statistically significantly different between the three tested costal levels (level 3 vs. level 2: level 4 vs. level 2: level 4 vs. level 3: t-values and p-values respectively t = - 0.36, p = 0.723: t = 0.35, p = 0.728: t = 0.71, p = 0.484). Conclusions: Vibration transmittance analysis differentiates the intact sternum, wire fixation with exact apposition, and wire fixation with a gap. The gap detection capability is not dependent on the tested costal level. The method may prove useful in the early detection of sternal instability and warrants further exploration.</p>
Kokoelmat
  • TUNICRIS-julkaisut [23497]
Kalevantie 5
PL 617
33014 Tampereen yliopisto
oa[@]tuni.fi | Tietosuoja | Saavutettavuusseloste
 

 

Selaa kokoelmaa

TekijätNimekkeetTiedekunta (2019 -)Tiedekunta (- 2018)Tutkinto-ohjelmat ja opintosuunnatAvainsanatJulkaisuajatKokoelmat

Omat tiedot

Kirjaudu sisäänRekisteröidy
Kalevantie 5
PL 617
33014 Tampereen yliopisto
oa[@]tuni.fi | Tietosuoja | Saavutettavuusseloste