A Fully Printed Ultra-Thin Charge Amplifier for On-Skin Biosignal Measurements

Laurila, Mika-Matti; Matsui, Hiroyuki; Shiwaku, R; Peltokangas, Mikko; Verho, Jarmo; Lozano Montero, Karem; Sekine, Tomohito; Vehkaoja, Antti; Oksala, Niku; Tokito, Shizuo; Mäntysalo, Matti (2019)

 
Avaa tiedosto
A_fully_printed_ultra_thin_charge_amplifier_for_on_skin_biosignal_measurements.pdf (871.1Kt)
Lataukset: 



Laurila, Mika-Matti
Matsui, Hiroyuki
Shiwaku, R
Peltokangas, Mikko
Verho, Jarmo
Lozano Montero, Karem
Sekine, Tomohito
Vehkaoja, Antti
Oksala, Niku
Tokito, Shizuo
Mäntysalo, Matti
2019

IEEE Journal of the Electron Devices Society
This publication is copyrighted. You may download, display and print it for Your own personal use. Commercial use is prohibited.
doi:10.1109/jeds.2019.2915028
Näytä kaikki kuvailutiedot
Julkaisun pysyvä osoite on
https://urn.fi/urn:nbn:fi:tuni-202001221465

Kuvaus

Peer reviewed
Tiivistelmä
In this contribution, we propose a fully printed charge amplifier for on-skin biosignal measurements. The amplifier is fabricated on an ultra-thin parylene substrate and consists of organic transistors, integrated bias and feedback resistors, and a feedback capacitor. The fabrication process utilizes inkjet-printed Ag ink for source, drain, gate, and capacitor electrode metallization as well as for the interconnects between the amplifier elements. Dispensed polystyrene, 2,7-dihexyl-dithieno[2,3-d:2',3'-d']benzo[1,2-b:4,5-b']dithiophene (PS:DTBDT-C6), is used as the transistor channel material, dispensed poly(3-hexylthiophene) (P3HT) as the high-resistivity material for the printed resistors, and parylene as the capacitor dielectric. A pass band optimized for pulse-wave measurement (60 mHz to 36 Hz) is achieved with a maximum charge amplification of 1.6 V/nC. To demonstrate the potential of the proposed printed amplifier, a radial arterial pulsewave signal recorded with a printed piezoelectric poly(vinylidenefluoride-co-trifluoroethylene) (PVDF-TrFE) sensor was fed to it and the output was analyzed to quantify the similarity of the pulse-wave features calculated from the original signal and the amplifier output. The amplified signal contains all the essential features of a pulse wave, such as both systolic waves, the dicrotic notch, and diastolic wave, which enable the accurate derivation of the clinically relevant indices utilized in the evaluation of vascular health.
Kokoelmat
  • TUNICRIS-julkaisut [24669]
Kalevantie 5
PL 617
33014 Tampereen yliopisto
oa[@]tuni.fi | Tietosuoja | Saavutettavuusseloste
 

 

Selaa kokoelmaa

TekijätNimekkeetTiedekunta (2019 -)Tiedekunta (- 2018)Tutkinto-ohjelmat ja opintosuunnatAvainsanatJulkaisuajatKokoelmat

Omat tiedot

Kirjaudu sisäänRekisteröidy
Kalevantie 5
PL 617
33014 Tampereen yliopisto
oa[@]tuni.fi | Tietosuoja | Saavutettavuusseloste