Detection of beat-to-beat intervals from wrist photoplethysmography in patients with sinus rhythm and atrial fibrillation after surgery

Tarniceriu, Adrian; Harju, Jarkko; Vehkaoja, Antti; Parak, Jakub; Delgado-Gonzalo, Ricard; Renevey, Philippe; Yli-Hankala, Arvi; Korhonen, Ilkka (2018)

 
Avaa tiedosto
Detection_of_Beat_to_Beat_Intervals_from_Wrist_Photoplethysmography_in_Patients_with_Sinus_Rhythm_and_Atrial_Fibrillation_after_Surgery_post_print.pdf (452.4Kt)
Lataukset: 



Tarniceriu, Adrian
Harju, Jarkko
Vehkaoja, Antti
Parak, Jakub
Delgado-Gonzalo, Ricard
Renevey, Philippe
Yli-Hankala, Arvi
Korhonen, Ilkka
2018

This publication is copyrighted. You may download, display and print it for Your own personal use. Commercial use is prohibited.
doi:10.1109/bhi.2018.8333387
Näytä kaikki kuvailutiedot
Julkaisun pysyvä osoite on
https://urn.fi/URN:NBN:fi:tty-201809252333

Kuvaus

Peer reviewed
Tiivistelmä
<p>Wrist photoplethysmography (PPG) allows unobtrusive monitoring of the heart rate (HR). PPG is affected by the capillary blood perfusion and the pumping function of the heart, which generally deteriorate with age and due to the presence of cardiac arrhythmia. The performance of wrist PPG in monitoring beat-to-beat HR in older patients with arrhythmia has not been reported earlier. We monitored PPG from wrist in 18 patients recovering from surgery in the post-anesthesia care unit, and evaluated the inter-beat interval (IBI) detection accuracy against ECG based R-to-R intervals (RRI). Nine subjects had sinus rhythm (SR, 68.0y ± 10.2y, 6 males) and nine subjects had atrial fibrillation (AF, 71.3y ± 7.8y, 4 males) during the recording. For the SR group, 99.44% of the beats were correctly identified, 2.39% extra beats were detected, and the mean absolute error (MAE) was 7.34 ms. For the AF group, 97.49% of the heartbeats were correctly identified, 2.26% extra beats were detected, and the MAE was 14.31 ms. IBI from the PPG were hence in close agreement with the ECG reference in both groups. The results suggest that wrist PPG provides a comfortable alternative to ECG during low motion and can be used for long-term monitoring and screening of AF episodes.</p>
Kokoelmat
  • TUNICRIS-julkaisut [20247]
Kalevantie 5
PL 617
33014 Tampereen yliopisto
oa[@]tuni.fi | Tietosuoja | Saavutettavuusseloste
 

 

Selaa kokoelmaa

TekijätNimekkeetTiedekunta (2019 -)Tiedekunta (- 2018)Tutkinto-ohjelmat ja opintosuunnatAvainsanatJulkaisuajatKokoelmat

Omat tiedot

Kirjaudu sisäänRekisteröidy
Kalevantie 5
PL 617
33014 Tampereen yliopisto
oa[@]tuni.fi | Tietosuoja | Saavutettavuusseloste