Extending the Faraday cup aerosol electrometer based calibration method up to 5 µm

Järvinen, A.; Keskinen, J.; Yli-Ojanperä, J. (2018-08-03)

 
Avaa tiedosto
J_rvinen_et_al_Calibration_rev_11.pdf (543.9Kt)
Lataukset: 



Järvinen, A.
Keskinen, J.
Yli-Ojanperä, J.
03.08.2018

Aerosol Science and Technology
This publication is copyrighted. You may download, display and print it for Your own personal use. Commercial use is prohibited.
doi:10.1080/02786826.2018.1472742
Näytä kaikki kuvailutiedot
Julkaisun pysyvä osoite on
https://urn.fi/URN:NBN:fi:tuni-201910033665

Kuvaus

Peer reviewed
Tiivistelmä
<p>A Faraday cup aerosol electrometer based electrical aerosol instrument calibration setup from nanometers up to micrometers has been designed, constructed, and characterized. The set-up utilizes singly charged seed particles, which are grown to the desired size by condensation of diethylhexyl sebacate. The calibration particle size is further selected with a Differential Mobility Analyzer (DMA). For micrometer sizes, a large DMA was designed, constructed, and characterized. The DMA electrical mobility resolution was found to be 7.95 for 20 L/min sheath and 2 L/min sample flows. The calibration is based on comparing the instrument’s response against the concentration measured with a reference Faraday cup aerosol electrometer. The set-up produces relatively high concentrations in the micrometer size range (more than 2500 1/cm<sup>3</sup> at 5.3 µm). A low bias flow mixing and splitting between the reference and the instrument was constructed from a modified, large-sized mixer and a four-port flow splitter. It was characterized at different flow rates and as a function of the particle size. Using two of the four outlet ports at equal 1.5 L/min flow rates, the particle concentration bias of the flow splitting was found to be less than ±1% in the size range of 3.6 nm–5.3 µm. The developed calibration set-up was used to define the detection efficiency of a condensation particle counter from 3.6 nm to 5.3 µm with an expanded measurement uncertainty (k = 2) of less than 4% over the entire size range and less than 2% for most of the measurement points.</p>
Kokoelmat
  • TUNICRIS-julkaisut [25675]
Kalevantie 5
PL 617
33014 Tampereen yliopisto
oa[@]tuni.fi | Tietosuoja | Saavutettavuusseloste
 

 

Selaa kokoelmaa

TekijätNimekkeetTiedekunta (2019 -)Tiedekunta (- 2018)Tutkinto-ohjelmat ja opintosuunnatAvainsanatJulkaisuajatKokoelmat

Omat tiedot

Kirjaudu sisäänRekisteröidy
Kalevantie 5
PL 617
33014 Tampereen yliopisto
oa[@]tuni.fi | Tietosuoja | Saavutettavuusseloste