Miniature Coplanar Implantable Antenna on Thin and Flexible Platform for Fully Wireless Intracranial Pressure Monitoring System

Khan, M. Waqas A.; Moradi, Elham; Sydänheimo, Lauri; Björninen, Toni; Rahmat-Samii, Yahya; Ukkonen, Leena (2016-12-14)

 
Avaa tiedosto
Khan_et_al._2016.pdf (2.534Mt)
Lataukset: 



Khan, M. Waqas A.
Moradi, Elham
Sydänheimo, Lauri
Björninen, Toni
Rahmat-Samii, Yahya
Ukkonen, Leena
14.12.2016

International Journal of Antennas and Propagation
9161083
doi:10.1155/2017/9161083
Näytä kaikki kuvailutiedot
Julkaisun pysyvä osoite on
https://urn.fi/URN:NBN:fi:tty-201701101039

Kuvaus

Peer reviewed
Tiivistelmä
Minimally invasive approach to intracranial pressure monitoring is desired for long-term diagnostics. The monitored pressure is transmitted outside the skull through an implant antenna. We present a new miniature (6 mm × 5 mm) co-planar implant antenna and its integration on a sensor platform to establish a far-field data link for the sensor readout at distances of 0.5-to-1 meters. The implant antenna was developed using full-wave electromagnetic simulator and measured in a liquid phantom mimicking the dielectric properties of the human head. It achieved impedance reflection coefficient better than −10 dB from 2.38 GHz to 2.54 GHz which covers the targeted industrial, scientific, and medical band. Experiments resulted into an acceptable peak gain of approximately −23 dBi. The implant antenna was submerged in the liquid phantom and interfaced to a 0.5 mW voltage controlled oscillator. To verify the implant antenna performance as a part of the ICP monitoring system, we recorded the radiated signal strength using a spectrum analyzer. Using a half-wavelength dipole as the receiving antenna, we captured approximately −58.7 dBm signal at a distance of 1 m from the implant antenna which is well above for the reader with sensitivity of -80 dBm.
Kokoelmat
  • TUNICRIS-julkaisut [22451]
Kalevantie 5
PL 617
33014 Tampereen yliopisto
oa[@]tuni.fi | Tietosuoja | Saavutettavuusseloste
 

 

Selaa kokoelmaa

TekijätNimekkeetTiedekunta (2019 -)Tiedekunta (- 2018)Tutkinto-ohjelmat ja opintosuunnatAvainsanatJulkaisuajatKokoelmat

Omat tiedot

Kirjaudu sisäänRekisteröidy
Kalevantie 5
PL 617
33014 Tampereen yliopisto
oa[@]tuni.fi | Tietosuoja | Saavutettavuusseloste