Evaluation of dry electrodes in canine heart rate monitoring

Virtanen, Juhani; Somppi, Sanni; Törnqvist, Heini; Jeyhani, Vala; Fiedler, Patrique; Gizatdinova, Yulia; Majaranta, Päivi; Väätäjä, Heli; Cardó, A.V.; Lekkala, Jukka; Tuukkanen, Sampo; Surakka, Veikko; Vainio, Outi; Vehkaoja, Antti; Kuosmanen, Yulia (2018)

 
Avaa tiedosto
sensors_18_01757_v2.pdf (4.501Mt)
Lataukset: 



Virtanen, Juhani
Somppi, Sanni
Törnqvist, Heini
Jeyhani, Vala
Fiedler, Patrique
Gizatdinova, Yulia
Majaranta, Päivi
Väätäjä, Heli
Cardó, A.V.
Lekkala, Jukka
Tuukkanen, Sampo
Surakka, Veikko
Vainio, Outi
Vehkaoja, Antti
Kuosmanen, Yulia
2018

Sensors
1757
doi:10.3390/s18061757
Näytä kaikki kuvailutiedot
Julkaisun pysyvä osoite on
https://urn.fi/urn:nbn:fi:tty-201806212014Julkaisun pysyvä osoite on
https://urn.fi/urn:nbn:fi:uta-201806132002

Kuvaus

Peer reviewed
Tiivistelmä
<p>The functionality of three dry electrocardiogram electrode constructions was evaluated by measuring canine heart rate during four different behaviors: Standing, sitting, lying and walking. The testing was repeated (n = 9) in each of the 36 scenarios with three dogs. Two of the electrodes were constructed with spring-loaded test pins while the third electrode was a molded polymer electrode with Ag/AgCl coating. During the measurement, a specifically designed harness was used to attach the electrodes to the dogs. The performance of the electrodes was evaluated and compared in terms of heartbeat detection coverage. The effect on the respective heart rate coverage was studied by computing the heart rate coverage from the measured electrocardiogram signal using a pattern-matching algorithm to extract the R-peaks and further the beat-to-beat heart rate. The results show that the overall coverage ratios regarding the electrodes varied between 45-95% in four different activity modes. The lowest coverage was for lying and walking and the highest was for standing and sitting.</p>
Kokoelmat
  • TUNICRIS-julkaisut [20247]
Kalevantie 5
PL 617
33014 Tampereen yliopisto
oa[@]tuni.fi | Tietosuoja | Saavutettavuusseloste
 

 

Selaa kokoelmaa

TekijätNimekkeetTiedekunta (2019 -)Tiedekunta (- 2018)Tutkinto-ohjelmat ja opintosuunnatAvainsanatJulkaisuajatKokoelmat

Omat tiedot

Kirjaudu sisäänRekisteröidy
Kalevantie 5
PL 617
33014 Tampereen yliopisto
oa[@]tuni.fi | Tietosuoja | Saavutettavuusseloste