Wrist-worn device combining PPG and ECG can be reliably used for atrial fibrillation detection in an outpatient setting
Saarinen, Harri Juhani; Joutsen, Atte; Korpi, Kirsi; Halkola, Tuomas; Nurmi, Marko; Hernesniemi, Jussi; Vehkaoja, Antti (2023-02-09)
Saarinen, Harri Juhani
Joutsen, Atte
Korpi, Kirsi
Halkola, Tuomas
Nurmi, Marko
Hernesniemi, Jussi
Vehkaoja, Antti
09.02.2023
Frontiers in Cardiovascular Medicine
1100127
Julkaisun pysyvä osoite on
https://urn.fi/URN:NBN:fi:tuni-202303303337
https://urn.fi/URN:NBN:fi:tuni-202303303337
Kuvaus
Peer reviewed
Tiivistelmä
<p>Aims: The aim was to validate the performance of a monitoring system consisting of a wrist-worn device and a data management cloud service intended to be used by medical professionals in detecting atrial fibrillation (AF). Methods: Thirty adult patients diagnosed with AF alone or AF with concomitant flutter were recruited. Continuous photoplethysmogram (PPG) and intermittent 30 s Lead I electrocardiogram (ECG) recordings were collected over 48 h. The ECG was measured four times a day at prescheduled times, when notified due to irregular rhythm detected by PPG, and when self-initiated based on symptoms. Three-channel Holter ECG was used as the reference. Results: The subjects recorded a total of 1,415 h of continuous PPG data and 3.8 h of intermittent ECG data over the study period. The PPG data were analyzed by the system’s algorithm in 5-min segments. The segments containing adequate amounts, at least ~30 s, of adequate quality PPG data for rhythm assessment algorithm, were included. After rejecting 46% of the 5-min segments, the remaining data were compared with annotated Holter ECG yielding AF detection sensitivity and specificity of 95.6 and 99.2%, respectively. The ECG analysis algorithm labeled 10% of the 30-s ECG records as inadequate quality and these were excluded from the analysis. The ECG AF detection sensitivity and specificity were 97.7 and 89.8%, respectively. The usability of the system was found to be good by both the study subjects and the participating cardiologists. Conclusion: The system comprising of a wrist device and a data management service was validated to be suitable for use in patient monitoring and in the detection of AF in an ambulatory setting. Clinical Trial Registration: ClinicalTrials.gov/, NCT05008601.</p>
Kokoelmat
- TUNICRIS-julkaisut [20250]