Dosimetric effect of rotational setup errors in volumetric modulated arc therapy and field-in-field treatment of left-sided breast cancer
Heikkilä, Annele; Boman, Eeva; Rossi, Maija; Vanhanen, Antti; Mankinen, Mikko; Postema, Michiel; Koivumäki, Tuomas (2024)
Heikkilä, Annele
Boman, Eeva
Rossi, Maija
Vanhanen, Antti
Mankinen, Mikko
Postema, Michiel
Koivumäki, Tuomas
2024
103203
Julkaisun pysyvä osoite on
https://urn.fi/URN:NBN:fi:tuni-202405145860
https://urn.fi/URN:NBN:fi:tuni-202405145860
Kuvaus
Peer reviewed
Tiivistelmä
Setup errors are an important factor in the dosimetric accuracy of radiotherapy delivery. In this study, we investigated how rotational setup errors influence the dose distribution in volumetric modulated arc therapy (VMAT) and tangential field-in-field (FiF) treatment of left-sided breast cancer with supraclavicular lymph node involvement in deep inspiration breath hold. Treatment planning computed tomography images and radiotherapy plans of 20 patients were collected retrospectively for the study. Rotational setup errors up to 3° were simulated by rotating the planning images, and the resulting dosimetric changes were calculated. With rotational setup errors up to 3°, the median decrease of V95% to clinical target volume was less than 0.8 percentage point in both VMAT and FiF plans. The dose distribution of the heart and left anterior descending artery was more stable with respect to rotations in VMAT plans compared to FiF plans. Correction of 1° setup errors is recommended due to increased doses to the heart and left anterior descending artery after 1° setup errors.