Direct Measurement of Contraction Force in Human Cardiac Tissue Model Using Piezoelectric Cantilever Sensor Technique
Virtanen, J.; Toivanen, M.; Toimela, T.; Heinonen, T.; Tuukkanen, S. (2019)
Virtanen, J.
Toivanen, M.
Toimela, T.
Heinonen, T.
Tuukkanen, S.
2019
Julkaisun pysyvä osoite on
https://urn.fi/urn:nbn:fi:tuni-201911075814
https://urn.fi/urn:nbn:fi:tuni-201911075814
Kuvaus
Peer reviewed
Tiivistelmä
A proof-of-concept method for measuring cardiac tissue contraction force using an in-house-developed piezoelectric cantilever sensor system is demonstrated. Contracting forces of 7.2 to 16.6 μN were measured from a human cardiac tissue construct. Beating cardiac tissue constructs were monitored in-situ under a microscope during the contraction force measurements. Development of the measurement method allows very low forces such as the ones that appear in biological small scale systems to be determined.